加州分析仪器公司( CAI)的总部设在奥伦治,加州一直是优质的气体分析仪和系统的领先供应商,用于工业,环境,工艺,健康和安全,以及超过25年的汽车排放测量应用。 CAI的目标是提供创新,成本效益和可靠的解决方案。我们相信这是我们的客户对产品的要求和支持的需求不可替代的个人关注。 CAI利用技术,如化学发光法,非色散红外( NDIR ) ,火焰离子化检测器( FID ) ,顺磁,红外光谱,红外光声光谱( PAS )测量CO , CO2,NO , NO2 , N2O , NH3 , O2 , CH4 , SF6 , SO2 , HCL气体。 CAI的计算机控制分析仪目前已安装在100多个国家的数千设施 
  • 产品特点
  • 技术参数
  • 资料下载
实时在线读数
快速响应
%,PPM to PPb
低维护
低成本使用
直观的用户界面
可移动
SPECIFICATIONS
DETECTOR: Chemiluminescence (CLD) Photodiode
(thermally stabilized with Peltier Cooler)
NO/NOx RANGES: 5 to 5,000 ppm NO/NOX
(Four user programmable ranges)
RESPONSE TIME: Typically < 2 seconds to 90% Full Scale
REPEATABILITY: Better than 0.5% of Full Scale
LINEARITY: Better than 0.5% of Full Scale
NOISE: Typically less than 1% of Full Scale
ZERO & SPAN DRIFT: Less than 1% of Full Scale per 24 Hours
ZERO & SPAN ADJUSTMENT: Via front panel, TCP/IP or RS-232
NH3 , HCN & SO2 EFFECT: Not detectable with 100 ppm
CO2 EFFECT: Less than 2.0% with 10% CO2
H2O EFFECT: Less than 0.8% total with chiller
FLOW CONTROL: Electronic Proportional Pressure Controller
SAMPLE FLOW RATE: Typically 1.5 to 2.5 LPM (0.6 LPM with low
flow option)
CONVERTER: Vitreous Carbon Material @ 205°C >95% efficiency
OZONATOR: Ultraviolet Lamp
AIR OR O2 REQUIREMENTS: Less than 0.01 ppm
NOX at 240 cc/Min. @ 25 psig (Dew Point < -10ºC)
NO/NOx CONTROL: Manual/Remote/Auto Cycle
OUTPUTS AVAILABLE: TCP/IP, RS232,
Four Scalable Analog 0-10 V / 4-20 mA
DISCRETE ALARMS: (Local & Remote Adjustable)
General Fault/ TTL Logic (Ground True)
Calibration Failure/ TTL Logic (Ground True)
High Concentration (2 each)/ TTL Logic (Ground True)
DIGITAL DIAGNOSTICS: Control Voltages, Temperatures,
Pressures, Flow Parameters
KEYPAD DISPLAYS: Factory Settings, TCP/IP Address,
Passwords (4), Scalable Analog Output Voltages,
Full Scale Range Select, Auto Cal Times
SPECIAL FEATURES: Calculated NO2 derived from NOX converter
efficiency, Auto Ranging, Auto Calibration (adjustable through internal
clock) Less than 3 cc Gold Plated Reaction Chamber
DISPLAY: 3” x 5” Back lit LCD
SAMPLE TEMPERATURE: Up to 50°C Noncondensing on CLD;
85-100oC on HCLD
OVEN TEMPERATURE: 85oC on HCLD; up to 100oC on request
AMBIENT TEMPERATURE: 5 to 40°C
AMBIENT HUMIDITY: Less than 90% RH Noncondensing
WARM-UP TIME: 1 Hour (Typical)
FITTINGS: 1/4 Inch Tube
POWER REQUIREMENTS: 115/230 (±10%) VAC; 50/60 Hz;
200 Watts (350 watts with pump)
DIMENSIONS: 5¼ H x 19 W x 23 D (Inches)
Weight: 45 lbs.